Researchers validate simplified lateral force calibration technique for atomic force microscopy

Mar 29, 2012 By Rachel Cannara
In the diamagnetic lateral force calibration method, an AFM cantilever (shown here is a NIST HammerHead cantilever) presses against the surface of a piece of graphite that levitates in a magnetic field. When the magnetic field is moved horizontally in the AFM, friction between the AFM tip and graphite surface causes the cantilever to twist. This twist leads to a change in the lateral signal in the AFM that is used to calibrate friction directly, based on the spring constant of the graphite in the magnetic field. 

(PhysOrg.com) -- Researchers from the NIST Center for Nanoscale Science and Technology and the NIST Material Measurement Laboratory have demonstrated that a simpler technique for calibrating lateral sensitivity in an atomic force microscope (AFM) agrees with an earlier method developed at NIST to within 5 %.

The equivalence of these two independent methods represents an important step towards traceable accuracy in lateral force microscopy and will allow scientists to better understand the atomic-scale origins of across a wide range of materials.

The “HammerHead” (HH) relies on precise positioning of the arms of a specially-fabricated, tee-shaped cantilever over well-defined alignment marks in a surface; a torque is applied at different locations on the cantilever arm by pressing it against a small sphere attached to the edge of the surface.

The ratio of the change in the normal (vertical) signal to the lateral signal can be used to calibrate sensitivity and extract friction forces corresponding to the lateral signals measured during an experiment.

The new “Diamagnetic Lateral Force Calibrator” (D-LFC) method, developed at Brown University, requires fewer independent measurements. The AFM cantilever presses against the surface of a piece of graphite that levitates in a magnetic field. When the magnetic field is moved horizontally in the AFM, the levitating graphite behaves like a mass on a very weak spring.

A lateral force is applied by the graphite to the tip of the AFM cantilever, causing the to twist. This twist leads to a change in the lateral signal in the AFM that can be used to calibrate friction directly, without the need for an independent measurement of the normal signal.

While the D-LFC method is preferable for most circumstances, because it uses fewer parameters and therefore has greater precision, the HH method can be advantageous if contact between the probe tip and the calibration surface must be avoided.

The researchers believe that the overall and comparability of these two methods establishes the importance of the D-LFC method as a valuable tool for unifying quantitative measurements of friction at the nanoscale, and establishes a potential path towards the development of lateral force standards.  

Explore further: Team finds electricity can be generated by dragging saltwater over graphene

More information: Quantitative comparison of two independent lateral force calibration techniques for the atomic force microscope, S. S. Barkley, et al., Review of Scientific Instruments 83, 023707 (2012). rsi.aip.org/resource/1/rsinak/v83/i2/p023707_s1

add to favorites email to friend print save as pdf

Related Stories

Scaling Friction Down to the Nano/Micro Realm

May 28, 2004

An improved method for correcting nano- and micro-scale friction measurements has been developed by researchers at the National Institute of Standards and Technology (NIST). The new technique should help designers ...

Shaking Reduces Friction

Jul 08, 2005

Lateral vibrations can control friction at the nanoscale, researchers reported in the 1 July 2005 issue of Physical Review Letters.

NIST imaging system maps nanomechanical properties

Dec 12, 2007

The National Institute of Standards and Technology has developed an imaging system that quickly maps the mechanical properties of materials—how stiff or stretchy they are, for example—at scales on the ...

Recommended for you

First direct observations of excitons in motion achieved

Apr 16, 2014

A quasiparticle called an exciton—responsible for the transfer of energy within devices such as solar cells, LEDs, and semiconductor circuits—has been understood theoretically for decades. But exciton ...

User comments : 0

More news stories

Thinnest feasible nano-membrane produced

A new nano-membrane made out of the 'super material' graphene is extremely light and breathable. Not only can this open the door to a new generation of functional waterproof clothing, but also to ultra-rapid filtration. The ...

Wiring up carbon-based electronics

Carbon-based nanostructures such as nanotubes, graphene sheets, and nanoribbons are unique building blocks showing versatile nanomechanical and nanoelectronic properties. These materials which are ordered ...

Scientists tether lionfish to Cayman reefs

Research done by U.S. scientists in the Cayman Islands suggests that native predators can be trained to gobble up invasive lionfish that colonize regional reefs and voraciously prey on juvenile marine creatures.

Leeches help save woman's ear after pit bull mauling

(HealthDay)—A pit bull attack in July 2013 left a 19-year-old woman with her left ear ripped from her head, leaving an open wound. After preserving the ear, the surgical team started with a reconnection ...

Six Nepalese dead, six missing in Everest avalanche

At least six Nepalese climbing guides have been killed and six others are missing after an avalanche struck Mount Everest early Friday in one of the deadliest accidents on the world's highest peak, officials ...