Fizeau interferometers for surfaces with different reflectivity

Dec 18, 2007
Intensities
Intensities of the interferences of common Fizeau interferometers, in this example calculated for a reference surface with a reflectivity of 0.5 and test sample surfaces with varying reflectivity between 0 and 100% and a variation of the distance d between test sample and reference piece over a wavelength (λ = 633 nm). The new method avoids the strong dependence of the intensity profile and the contrast on the reflectivity of the test sample's surface. Credit: PTB

Due to their stable design, Fizeau interferometers are used to determine the topography of surfaces such as, e.g., plane surfaces. PTB (Physikalisch-Technische Bundesanstalt, Germany) has now developed a method which makes it possible to analyse surfaces with different reflectivities in a simple way. This method can also be extended to dynamic measurements.

Fizeau interferometers generate an interference between the surface of a test sample and a reference surface that is brought close to the test sample. The interference image is recorded and analysed by an imaging optics system. The contrast and the shape of the interference signals depend, however, on the reflectivity of the test samples. The time and effort required for measuring and analysing the topographies of differently reflecting test pieces is therefore significantly increased.

The method of separating the wavefronts of the reference surface and the test sample surface in the plane of the reference surface – for which a patent has been applied for – uses a new generation of commercially available beam splitters, also called "on-axis beam splitters", which cause a separation of the polarisation directions of the incident light along the optical axis.

In comparison with common Fizeau interferometers, this has various advantages. On the one hand, through the generation of polarised light, the measurement can be traced back to a pure two-beam interference. The analysis of the signal is thus significantly facilitated and improved. Classic Fizeau interferometers, however, are based on the analysis of the multiple-beam interference.

On the other hand, by varying the direction of polarisation, a maximum contrast can be set, independent of the reflectivity of the test sample. In contrast to this, in conventional Fizeau interferometers, different reference surfaces with adapted reflectivity must be held available in case the reflectivity of the test samples varies strongly.

In order to increase the accuracy of analysis of common Fizeau interferometers, a variable phase is generated by varying mechanically the distance between the test sample and the reference surface (phase-shifter interferometry). Another advantage of the new method is that such a phase shifting becomes possible through the use of electro-optical components and thus without using mobile parts. Thanks to the increased measuring dynamics achieved in this way, it is, for example, possible to carry out topographical measurements in environments which are subjected to vibrational strain.

The new method allows a facilitated and improved analysis of the measurement data, requires only one calibrated reference surface instead of several, and opens up new possibilities of application in the field of dynamic interferometry. It is particularly suited for measurements on structured or unstructured surfaces with different reflectivities – especially in the optical and semiconductor industry.

Source: Physikalisch-Technische Bundesanstalt (PTB)

Explore further: Used MRI magnets get a second chance at life in high-energy physics experiments

Related Stories

A conversation with astronomer Dimitri Mawet

May 18, 2015

Associate Professor of Astronomy Dimitri Mawet has joined Caltech from the Paranal Observatory in Chile, where he was a staff astronomer for the Very Large Telescope. After earning his PhD at the University ...

Telescope to seek dust where other Earths may lie

Jan 22, 2015

The NASA-funded Large Binocular Telescope Interferometer, or LBTI, has completed its first study of dust in the "habitable zone" around a star, opening a new door to finding planets like Earth. Dust is a ...

Progress on detecting glucose levels in saliva

Jun 03, 2014

Researchers from Brown University have developed a new biochip sensor that can selectively measure concentrations of glucose in a complex solution similar to human saliva. The advance is an important step ...

Recommended for you

SLAC gears up for dark matter hunt with LUX-ZEPLIN

May 21, 2015

Researchers have come a step closer to building one of the world's best dark matter detectors: The U.S. Department of Energy (DOE) recently signed off on the conceptual design of the proposed LUX-ZEPLIN (LZ) ...

First images of LHC collisions at 13 TeV

May 21, 2015

Last night, protons collided in the Large Hadron Collider (LHC) at the record-breaking energy of 13 TeV for the first time. These test collisions were to set up systems that protect the machine and detectors ...

User comments : 0

Please sign in to add a comment. Registration is free, and takes less than a minute. Read more

Click here to reset your password.
Sign in to get notified via email when new comments are made.